 Name of Standard: National standard for low gauge pressure, negative gauge pressure and differential pressure in gaseous media
Name of Standard: National standard for low gauge pressure, negative gauge pressure and differential pressure in gaseous media
Code designation: ECM 170-5/02-022
Year of publication: 2002
Department: nr. 6013 CMI RI Brno
Guarantor: Ing. Zdeněk Krajíček, Ph.D., RNDr. Jiří Tesař, Ph.D (2002-2006)
Number of CMC lines provided: 2
| Mode | Range | Uncertainty (k = 2) | 
| Gauge pressure | 1 Pa – 3,2 kPa | 0,02 Pa + 8∙10-5∙p | 
| Negative gauge pressure | 1 Pa – 3,2 kPa | 0,02 Pa + 8∙10-5∙p | 
(p is the value of the generated pressure in pascals)
CMC magnitudes are expressed in terms of expanded uncertainty using an expansion factor of k = 2 (coverage probability of approximately 95%).
where stands for,
| p .......….... | pressure, | 
| kN ........…. | sensitivity constant of the instrument, | 
| N .....…..... | comparator indication during measurement, | 
| NK .....….... | comparator indication during zeroing, | 
| T.....…..... | temperature of the piston-cylinder, | 
| α......…...... | coefficient of thermal expansion of the gauge material, | 
| g ......…..... | local acceleration due to gravity, | 
| gN …….… | normal acceleration due to gravity, | 
| T0 .............. | reference temperature, | 
| ρa .............. | the current density of the atmosphere, | 
| ρN ............. | the density of the atmosphere under normal conditions, | 
| ρm .............. | density of the calibration weight. | 
Etalon has successfully participated in two published international comparisons.
 
                    