Calibration and measurement capabilities

thermometry and humidity
Interní identifikátor Quantity Calibration and measurement capabilities Measurement conditions Note Technical competence Link
1012-311 teplota
  • 419.527 °C - 419.527 °C

     

    0.12 K

  • temperature controlled 3-zone furnace 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA
1012-312 teplota
  • 660.323 °C - 660.323 °C

     

    0.12 K

  • heat pipe 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA
1012-313 teplota
  • 961.78 °C - 961.78 °C

     

    0.17 K

  • heat pipe 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA
teplota
  • 0 °C - 220 °C

     

    0.4 K

  • temperature controlled water bath and oil bath 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA
teplota
  • 220 °C - 550 °C

     

    0.5 K

  • temperature controlled oil bath and salt bath 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA
teplota
  • 550 °C - 1100 °C

     

    0.8 K

  • temperature controlled salt bath and high temperature furnace 
inhomogeneity uncertainty must be added to the combined uncertainty quoted in the calibration report CIPM MRA