Field of measurement: electricity and magnetism
Quantity: Soft magnetic sheet and powder materials: RMS value of magnetic field strength
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
MAT10 (0.03 - 10) kA/m (5 - 5) 0.001
  • frequency 50 Hz
  • polarisation 0,1 T to 1,9 T
817-MP-C604
Quantity: Soft magnetic sheet and powder materials: specific apparent power
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
MAT11 (1 - 100) VA/kg (30 - 30) 0.001
  • frequency 50 Hz
  • polarisation 0,3 T to 1,7 T
817-MP-C604
Quantity: Soft magnetic bulk materials: magnetic saturation polarization
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
MAT20 (0.2 - 5) Am2/kg (15 - 15) 0.001 817-MP-C606
Quantity: Hard magnetic materials: remanent magnetic flux density
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
MAT23 (0.1 - 1.3) T (15 - 15) 0.001 817-MP-C605
Quantity: Hard magnetic materials: coercive field strength (HCB, HCJ)
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
MAT24 (30 - 400) kA/m (15 - 15) 0.001 817-MP-C605
Quantity: Hard magnetic material: maximum energy product (B·H)max
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
MAT25 (100 - 500) kJ/m3 (25 - 25) 0.001 817-MP-C605
Quantity: AC current
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
9.20 (1 - 10) kA (0.021 - 0.1) % 817-MP-C705
9.20 (1 - 10) kA (5 - 20) ꞌ 817-MP-C705
9.20 (1 - 30) kA (0.021 - 0.1) % 817-MP-C705
9.20 (1 - 30) kA (5 - 20) ꞌ 817-MP-C705
Quantity: magnetic flux density
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
(1 - 2000) mT (0.21 - 0.21) % 817-MP-C607
Quantity: electromagnetic fields above 50 kHz: AC magnetic flux density
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EMF14 (0.001 - 0.08) mT (20 - 20) 0.001
  • frequency 50 kHz to 100 kHz
817-MP-C608
Field of measurement: photometry and radiometry
Quantity: Luminous intensity
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(1 - 1.0E+4) cd (0.9 - 0.8) %
  • Correlated colour temperature (2000 up to 2856) K
818-MP-C811
(1 - 1.0E+4) cd (0.8 - 0.9) %
  • Correlated color temperature (2856 up to 3200) K
818-MP-C811
17 (1 - 2.0E+4) cd (0.8 - 0.8) % 818-MP-C811
Quantity: Illuminance responsivity, tungsten source
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(- - -) A/lx (0.9 - 0.8) %
  • Illuminance (0.1 up to 10) lx
  • Correlated colour temperature 2856 K
818-MP-C811
(- - -) A/lx (0.8 - 0.8) %
  • Illuminance (10 up to 1000) lx
  • Correlated colour temperature 2856 K
818-MP-C811
(- - -) A/lx (0,8 - 1,2) %
  • Illuminance (1000 up to 30000) lx
  • Correlated colour temperature 2856 K
818-MP-C811
Quantity: Luminous flux
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(10 - 3.0E+4) lm (0,9 - 1,5) %
  • Correlated colour temperature (2000 up to 3200) K
818-MP-C807
19 (10 - 2.0E+4) lm (1 - 1) % 818-MP-C807
Quantity: Illuminance
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0.1 - 10) lx (0,9 - 0,8) %
  • Correlated colour temperature (2000 up to 3200) K
818-MP-C811
(10 - 1000) lx (0,8 - 0,8) %
  • Correlated colour temperature (2000 up to 3200) K
(1000 - 3.0E+4) lx (0,8 - 1,2) %
  • Correlated colour temperature (2000 up to 3200) K
818-MP-C811
6 (5 - 5.0E+4) lx (2.2 - 2.2) % 818-MP-C801
6 (1 - 5.0E+4) lx (0.8 - 0.8) % 818-MP-C811
Quantity: Transmittance, regular spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0.65 - 1) (0,0008 - 0,0008)
  • Wavelength (380 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0041 - 0,0041)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0013 - 0,0013)
  • Wavelength (400 up to 700) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0017 - 0,0017)
  • Wavelength (700 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,0015 - 0,0015)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,00022 - 0,00022)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Absorbance, regular, spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0 - 0.19) (0.0003 - 5.0E-5)
  • Wavelength (380 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0019 - 0.0052)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0009 - 0.0019)
  • Wavelength (400 up to 700) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0011 - 0.0025)
  • Wavelength (700 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0022 - 0.0339)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0003 - 0.0048)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.03 - 0.0969)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.0017 - 0.0362)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Wavelength
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(400 - 1650) nm (2.0E-6 - 2.0E-6) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
11 (200 - 1700) nm (0.2 - 0.2) nm 818-MP-C810
11 (200 - 1700) nm (0.2 - 0.2) nm 818-MP-C810
16 (800 - 1650) nm (4.0E-9 - 4.0E-9) nm 818-MP-C813
22 (365 - 923) nm (0.4 - 0.4) nm 818-MP-C806
Quantity: Spectral reflectance factor R
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
1 (0 - 100) % ((0,1 + 0,010R) - (0,1 + 0,010R)) % (abs.)
  • Wavelength (380 up to 460) nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,12 + 0,008R) - (0,12 + 0,008R)) % (abs.)
  • Wavelength 465 up to 780 nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,17 + 0,014R) - (0,17 + 0,014R)) % (abs.)
  • Wavelength (380 up to 780) nm
  • geometry 0º / 45º
818-MP-C802
1 (0 - 100) % ((0,1 + 0,010R) - (0,1 + 0,010R)) % (abs.)
  • Wavelength (380 up to 460) nm
  • geometry 8º / t, 8º / d
818-MP-C802